FDA Adverse Event Injury Summary report: N

ADAPTA DR

MDR report key: 4161075 · Received October 10, 2014

Report

Report Number
9614453-2014-02331
Event Type
Injury
Date Received
October 10, 2014
Date of Event
July 22, 2014
Report Date
July 22, 2014
Manufacturer
IPG MFG SWITZERLAND
Product Code
NVZ
PMA / PMN Number
P980035
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US WHERE A SIMILAR MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. PRODUCT EVENT SUMMARY: THE DEVICE WAS RETURNED AND ANALYZED. ANALYSIS REVEALED TANTALUM CAPACITOR SHORTING LOW RESISTANCE. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THEY WERE UNABLE TO INTERROGATE THE IMPLANTABLE PULSE GENERATOR (IPG). THE DEVICE DID NOT RESPOND TO ANY INTERROGATION OR MAGNET. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
642458 ADAPTA DR PULSE GENERATOR, PERMANENT, IMPLANTABLE NVZ IPG MFG SWITZERLAND ADDR03

Patients

Seq Age Sex Outcome Treatment
1 00094 YR Hospitalization| R