FDA Adverse Event
Injury
Summary report: N
ADAPTA DR
MDR report key: 4161075
·
Received October 10, 2014
Report
- Report Number
- 9614453-2014-02331
- Event Type
- Injury
- Date Received
- October 10, 2014
- Date of Event
- July 22, 2014
- Report Date
- July 22, 2014
- Manufacturer
- IPG MFG SWITZERLAND
- Product Code
- NVZ
- PMA / PMN Number
- P980035
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THIS EVENT OCCURRED OUTSIDE THE US WHERE A SIMILAR MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. PRODUCT EVENT SUMMARY: THE DEVICE WAS RETURNED AND ANALYZED. ANALYSIS REVEALED TANTALUM CAPACITOR SHORTING LOW RESISTANCE. (B)(4).
Description of Event or Problem · 1
IT WAS REPORTED THAT THEY WERE UNABLE TO INTERROGATE THE IMPLANTABLE PULSE GENERATOR (IPG). THE DEVICE DID NOT RESPOND TO ANY INTERROGATION OR MAGNET. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 642458 | ADAPTA DR | PULSE GENERATOR, PERMANENT, IMPLANTABLE | NVZ | IPG MFG SWITZERLAND | ADDR03 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00094 YR | Hospitalization| R |