FDA Adverse Event
Injury
Summary report: N
PROTECTA CRT-D
MDR report key: 4161058
·
Received October 10, 2014
Report
- Report Number
- 9614453-2014-02329
- Event Type
- Injury
- Date Received
- October 10, 2014
- Date of Event
- July 2, 2014
- Report Date
- July 2, 2014
- Manufacturer
- IPG MFG SWITZERLAND
- Product Code
- NIK
- PMA / PMN Number
- P010031
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
PRODUCT EVENT SUMMARY: THE DEVICE WAS RETURNED AND ANALYZED. ANALYSIS WAS PERFORMED AND NO ANOMALIES WERE FOUND. THE RETURNED DEVICE INDICATED A DAMAGED GROMMET. THE RETURNED DEVICE INDICATED A LOOSE/DETACHED SET SCREW. THIS MODEL NUMBER IS NOT APPROVED FOR DISTRIBUTION IN THE UNITED STATES, HOWEVER, IT IS SIMILAR TO A DEVICE MARKETED IN THE U.S. THE EVENT IS BEING REPORTED DUE TO AN ALLEGED MALFUNCTION. (B)(4).
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING A LEAD REVISION PROCEDURE, AFTER THE LEAD WAS DISCONNECTED FROM THE DEVICE THE SETSCREW OF THE RV CHANNEL WAS DAMAGED. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 641254 | PROTECTA CRT-D | DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO | NIK | IPG MFG SWITZERLAND | D364TRG |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00045 YR | Hospitalization| R |