FDA Adverse Event Injury Summary report: N

INSYNC SENTRY

MDR report key: 4161052 · Received October 10, 2014

Report

Report Number
9614453-2014-02325
Event Type
Injury
Date Received
October 10, 2014
Date of Event
July 8, 2014
Report Date
July 8, 2014
Manufacturer
IPG MFG SWITZERLAND
Product Code
NIK
PMA / PMN Number
P010031
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SN
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. THIS MODEL NUMBER IS NOT APPROVED FOR DISTRIBUTION IN THE UNITED STATES; HOWEVER, IT IS SAME/SIMILAR TO A DEVICE MARKETED IN THE U.S. THIS EVENT OCCURRED OUTSIDE THE US AND PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. PRODUCT EVENT SUMMARY: THE DEVICE WAS RETURNED AND ANALYZED. ANALYSIS REVEALED THE RETURNED DEVICE INDICATED NORMAL BATTERY DEPLETION. ANALYSIS OF THE DEVICE MEMORY WAS PERFORMED AND NO ANOMALIES WERE FOUND. ELECTIVE REPLACEMENT INDICATOR (ERI) NOTED ON (B)(6) 2013. CONCOMITANT MEDICAL PRODUCTS: 693565 LEAD IMPLANTED: (B)(6) 2011. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) WAS EXPLANTED AND REPLACED DUE TO A PREMATURE END OF LIFE (EOL) INDICATOR. IN ADDITION, THE RIGHT VENTRICULAR (RV) LEAD AND LEFT VENTRICULAR (LV) LEAD EXHIBITED HIGH THRESHOLDS. AS A RESULT, THE RV AND LV LEADS WERE REPROGRAMMED TO HIGH OUTPUT, AND REMAIN IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
642168 INSYNC SENTRY DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO NIK IPG MFG SWITZERLAND 7298

Patients

Seq Age Sex Outcome Treatment
1 00034 YR 4196 LEAD