FDA Adverse Event Injury Summary report: N

SPRINT QUATTRO

MDR report key: 4161047 · Received October 10, 2014

Report

Report Number
2649622-2014-11668
Event Type
Injury
Date Received
October 10, 2014
Date of Event
July 1, 2014
Report Date
July 1, 2014
Manufacturer
MPRI
Product Code
LWS
PMA / PMN Number
P920015
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
ND, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

EVALUATION SUMMARY: ANALYSIS OF THE DEVICE MEMORY INDICATED OVERSENSING DUE TO NON-PHYSIOLOGIC SIGNALS/SIC (SENSING INTEGRITY COUNTER), AND THAT THE CRITERIA FOR THE RIGHT VENTRICULAR LEAD INTEGRITY ALERT WERE MET.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD HAD OVERSENSING AND NOISE DUE TO A SUSPECTED LEAD FRACTURE. INTERRUPTION OF THE SUPERIOR VENA CAVA (SVC) COIL AND COIL MIGRATION INTO THE RIGHT SUBCLAVIAN POCKET AREA WERE NOTED VIA FLUOROSCOPY AT TIME OF REVISION. THE RV LEAD WAS CAPPED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
642536 SPRINT QUATTRO DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS MPRI 6944-58

Patients

Seq Age Sex Outcome Treatment
1 00032 YR Hospitalization| R DVBB1D1 ICD