FDA Adverse Event Injury Summary report: N

CAPSUREFIX NOVUS

MDR report key: 4161042 · Received October 10, 2014

Report

Report Number
2649622-2014-11664
Event Type
Injury
Date Received
October 10, 2014
Date of Event
July 9, 2014
Report Date
July 17, 2014
Manufacturer
MPRI
Product Code
DTB
PMA / PMN Number
P930039
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US WHERE A SIMILAR MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED AND ANALYZED. ANALYSIS REVEALED THE PROXIMAL CONDUCTOR OF THE LEAD DEVELOPED A FRACTURE DUE TO FLEXING WHILE IN VIVO. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE IMPEDANCE ON THE ATRIAL LEAD WAS HIGH. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
642464 CAPSUREFIX NOVUS ELECTRODE, PACEMAKER, PERMANENT DTB MPRI 407652

Patients

Seq Age Sex Outcome Treatment
1 00065 YR Hospitalization| R