FDA Adverse Event
Malfunction
Summary report: N
CARELINK
MDR report key: 4161037
·
Received October 10, 2014
Report
- Report Number
- 2182208-2014-02759
- Event Type
- Malfunction
- Date Received
- October 10, 2014
- Date of Event
- July 24, 2014
- Report Date
- July 24, 2014
- Manufacturer
- MEDTRONIC, INC.
- Product Code
- KRG
- PMA / PMN Number
- P890003
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCT EVENT SUMMARY: ANALYSIS CONFIRMED THE REPORTED EVENT, THE PROGRAMMER HAD THE GRAY SCREEN UPON POWER UP THEN AN ERROR WITH INTERMITTENT BLANK SCREEN WITH JUST A FLASHING CURSOR. AS A RESULT THE HARD DRIVE WAS REPLACED, REIMAGED AND SOFTWARE WAS RELOADED. IT WAS ALSO NOTED THAT THE SYSTEM FAN WAS NOISY. (B)(4).
Description of Event or Problem · 1
IT WAS REPORTED THAT THE PROGRAMMER DISPLAYED AN ERROR MESSAGE WHEN POWERING UP. THE USER CYCLED POWER AND PROGRAMMER JUST SAT AT A BLACK SCREEN. THE PROGRAMMER HAS BEEN RETURNED FOR REPAIR. THERE WAS NO PATIENT INVOLVEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 642084 | CARELINK | PROGRAMMER, PACEMAKER | KRG | MEDTRONIC, INC. | 2090 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 2067 RADIOFREQUENCY (RF) HEAD |