CAPSUREFIX NOVUS

MDR report key: 4161012 · Received October 10, 2014

Report

Report Number
2649622-2014-11652
Event Type
Malfunction
Date Received
October 10, 2014
Date of Event
July 23, 2014
Report Date
July 23, 2014
Manufacturer
MPRI
Product Code
DTB
PMA / PMN Number
P930039
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED. ANALYSIS WAS PERFORMED AND NO ANOMALIES WERE FOUND. HELIX EXTENSION/RETRACTION AND LENGTH TESTING WERE ALL PERFORMED AND ALL RESULTS, INCLUDING ELECTRICAL TESTING, WERE WITHIN SPECIFIED PARAMETERS. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A LEAD REPLACEMENT PROCEDURE, THE REPLACEMENT LEAD¿S HELIX COULD NOT BE ADEQUATELY RETRACTED. THE LEAD WAS NOT IMPLANTED AND ANOTHER LEAD WAS USED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
642510 CAPSUREFIX NOVUS ELECTRODE, PACEMAKER, PERMANENT DTB MPRI 5076-52

Patients

Seq Age Sex Outcome Treatment
1 00071 YR 4194-78 LEAD