FDA Adverse Event Injury Summary report: N

INSYNC III

MDR report key: 4160987 · Received October 10, 2014

Report

Report Number
9614453-2014-02340
Event Type
Injury
Date Received
October 10, 2014
Date of Event
March 7, 2014
Report Date
July 23, 2014
Manufacturer
IPG MFG SWITZERLAND
Product Code
LWP
PMA / PMN Number
P010015
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NL
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. PRODUCT EVENT SUMMARY: THE DEVICE WAS RETURNED AND ANALYZED. ANALYSIS OF THE DEVICE REVEALED NORMAL BATTERY DEPLETION. (B)(4).

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: PERFORMANCE DATA COLLECTED FROM THE DEVICE WAS RECEIVED AND ANALYZED. ANALYSIS OF THE DEVICE MEMORY SHOWED THE BATTERY INDICATOR SIGNIFYING THAT IT IS TIME FOR DEVICE REPLACEMENT. ERI TRIGGERED ON (B)(6) 2014.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE HAD REACHED ELECTIVE REPLACEMENT INDICATOR (ERI) FAST. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
641146 INSYNC III PULSE-GENERATOR, DUAL CHAMBER, IMPLANTABLE LWP IPG MFG SWITZERLAND 8042U

Patients

Seq Age Sex Outcome Treatment
1 00079 YR Hospitalization| R