Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCT EVENT SUMMARY: A PROXIMAL PORTION WAS RECEIVED MEASURING 13.5 CM, AND THE DISTAL PORTION WAS RECEIVED MEASURING 43.5 CM WERE RECEIVED AND ANALYZED. THE PROXIMAL CONDUCTOR OF THE LEAD DEVELOPED A FRACTURE DUE TO FLEXING WHILE IN VIVO. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT RIGHT VENTRICULAR (RV) LEAD SHOWED NOISE AND OVERSENSING WITH SHORT INTERVAL COUNTS. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
640783 SPRINT QUATTRO SECURE DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS MPRI 694758

Patients

Seq Age Sex Outcome Treatment
1 00057 YR Hospitalization| R D154VWC ICD