FDA Adverse Event Injury Summary report: N

ATLAS PLUS DR

MDR report key: 4160828 · Received January 13, 2014

Report

Report Number
2938836-2014-01860
Event Type
Injury
Date Received
January 13, 2014
Date of Event
November 2, 2012
Manufacturer
ST. JUDE MEDICAL INC. CRMD
Product Code
LWS
PMA / PMN Number
P910023
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ALL INFO PROVIDED BY MFR, NO MEDWATCH FORM WAS RECEIVED. THIS HISTORICAL COMPLAINT IS BEING FILED AS PART OF A RETROSPECTIVE REVIEW OF COMPLAINT FILES IN RESPONSE TO A RECENT FDA INSPECTION. THERE IS NO CHANGE TO THE ACTUAL PERFORMANCE OF THE PRODUCT AND THIS REPORT ONLY REPRESENTS AN ENHANCEMENT TO THE REPORTING CRITERIA GOING FORWARD. (B)(6).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE WHICH INAPPROPRIATELY SUPPLIED THERAPY. THE INAPPROPRIATE THERAPY OCCURRED WHEN THE PT WAS IN SUPRAVENTRICULAR TACHYCARDIA AND THE DEVICE DIAGNOSED VT. REPROGRAMMING CHANGES WERE RECOMMENDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
26691 ATLAS PLUS DR IMPLANTABLE CARDIOVERTER DEFIBRILLATOR LWS ST. JUDE MEDICAL INC. CRMD V-243

Patients

Seq Age Sex Outcome Treatment
1 85 YR