FDA Adverse Event Malfunction Summary report: N

CAPSUREFIX MRI SURESCAN

MDR report key: 4160655 · Received October 10, 2014

Report

Report Number
2649622-2014-11795
Event Type
Malfunction
Date Received
October 10, 2014
Date of Event
July 2, 2014
Report Date
July 2, 2014
Manufacturer
MPRI
Product Code
NVN
PMA / PMN Number
P090013
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED AND ANALYZED AND NO ANOMALIES WERE FOUND. THE DISTAL LV (LOW VOLTAGE) ELECTRODE OF THE LEAD WAS COVERED IN BLOOD AND BODY TISSUE/FIBROTIC GROWTH. VISUAL SUMMARY ANALYSIS OF THE LEAD INDICATED DAMAGE AT IMPLANT AND STRETCHING. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PHYSICIAN WAS UNABLE TO ADVANCE THE IMPLANTABLE PACING LEAD PAST A POINT BEFORE THE SUPERIOR VENA CAVA (SVC) DURING THE ATTEMPTED RIGHT ATRIAL (RA) IMPLANT PROCEDURE. WHEN THE LEAD WAS REMOVED AND EXAMINED, CARDIAC TISSUE WAS FOUND TO BE EMBEDDED IN THE LEAD HELIX WHICH COULD NOT BE COMPLETELY REMOVED. THE PHYSICIAN REQUESTED A NEW LEAD FOR COMPLETION OF THE IMPLANT PROCEDURE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
641323 CAPSUREFIX MRI SURESCAN DRUG ELUTING PERMANENT RIGHT VENTRICULAR (RV) OR RIGHT ATRIAL (RA) PACEMAKER ELE NVN MPRI 5086MRI52

Patients

Seq Age Sex Outcome Treatment
1 00073 YR