FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4160499 · Received September 17, 2014

Report

Report Number
2032227-2014-24088
Event Type
Malfunction
Date Received
September 17, 2014
Date of Event
August 10, 2014
Report Date
August 18, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

INSULIN PUMP RECEIVED WITH BLANK DISPLAY DUE TO CORRODED ELECTRONIC ASSEMBLY. THE MOTOR ASSEMBLY WAS FOUND CORRODED. UNABLE TO VERIFY BUTTON ERROR ALARMS DUE TO BLANK DISPLAY. INSULIN PUMP RECEIVED WITH CRACKED CASE AT DISPLAY WINDOW CORNERS AND BATTERY TUBE THREADS, SEVERE SCRATCHES ON DISPLAY WINDOW, MISSING END CAP STICKER AND WITH BROKEN RESERVOIR TUBE LIP.

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.THIS MDR RELATED TO THE (B)(4) MANUFACTURING SITE HAS BEEN ASSIGNED A MEDWATCH NUMBER FROM THE MEDTRONIC (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULIN PUMP ALARMED BUTTON ERROR. THE CUSTOMER'S BLOOD GLUCOSE WAS 372 MG/DL. THE CUSTOMER STATED THAT THEY WENT SWIMMING WITH THE INSULIN PUMP IN THEIR POCKET. THE INSULIN PUMP HAD BEEN COMPLETELY SUBMERGED IN THE LAKE WATER. THE DISCONTINUATION OF THE DEVICE WAS ADVISED. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
574656 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP OYC MEDTRONIC MINIMED MMT-723NAB

Patients

Seq Age Sex Outcome Treatment
1 19 YR