FDA Adverse Event Injury Summary report: N

ENRHYTHM DR

MDR report key: 4160441 · Received October 10, 2014

Report

Report Number
9614453-2014-02374
Event Type
Injury
Date Received
October 10, 2014
Date of Event
July 31, 2014
Report Date
August 8, 2014
Manufacturer
IPG MFG SWITZERLAND
Product Code
NVZ
PMA / PMN Number
P980035
Removal / Correction Number
Z-1440-2010
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE DEVICE WAS RETURNED AND ANALYZED. ANALYSIS OF THE DEVICE REVEALED NORMAL BATTERY DEPLETION. THIS DEVICE WAS INCLUDED IN THAT FIELD ACTION, BUT RETURNED PRODUCT TESTING FOUND THE DEVICE DID NOT PERFORM AS DESCRIBED IN THE FIELD ACTION.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT PRODUCT: 407652 IMPLANTABLE PACING LEAD, IMPLANTED: (B)(6) 2005. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT WAS ADMITTED FOR DIZZINESS POSSIBLY DUE TO DEVICE AT ELECTIVE REPLACEMENT INDICATOR (ERI). THE DEVICE WAS EXPLANTED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
638333 ENRHYTHM DR PULSE GENERATOR, PERMANENT, IMPLANTABLE NVZ IPG MFG SWITZERLAND P1501DR

Patients

Seq Age Sex Outcome Treatment
1 00084 YR Hospitalization| R 407645 IMPLANTABLE PACING LEAD