PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4160415 · Received September 20, 2014

Report

Report Number
2032227-2014-26227
Event Type
Malfunction
Date Received
September 20, 2014
Date of Event
August 21, 2014
Report Date
August 21, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
DE, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP WAS RECEIVED WITH STUCK MOTOR ERROR ALARM LOOP DURING BOLUS DELIVERY. THE MOTOR WAS TESTED OUTSIDE OF THE DEVICE AND IT PASSED. THE MOTOR MAY HAVE HAD AN INTERMITTENT FAILURE THAT WAS NOT DETECTED DURING TESTING. THE EXCESSIVE NO DELIVERY TEST AND OCCLUSION TEST WERE NOT PERFORMED, DUE TO DEVICE CONSTANTLY ALARMING MOTOR ERROR. MOTOR POSITION ENCODER ERROR ALARM WAS NOT VERIFIED DUE TO OVER WRITTEN PUMP HISTORY. THE DEVICE WAS RECEIVED WITH MINOR SCRATCHED DISPLAY WINDOW, CRACKED CASE AT DISPLAY WINDOW CORNERS, AND CRACKED RESERVOIR TUBE LIP.

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER RECEIVED A MOTOR ERROR ALARM AS WELL AS A MOTOR POSITION ENCODER ERROR ALERT. CUSTOMER'S BLOOD GLUCOSE LEVEL AT THE TIME 176 MG/DL. ADVISED INSULIN PUMP WOULD BE REPLACED. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
585719 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP OYC MEDTRONIC MINIMED MMT-723NAB

Patients

Seq Age Sex Outcome Treatment
1 57 YR