FDA Adverse Event Injury Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 4160156 · Received October 10, 2014

Report

Report Number
2649622-2014-12014
Event Type
Injury
Date Received
October 10, 2014
Date of Event
July 7, 2014
Report Date
July 8, 2014
Manufacturer
MPRI
Product Code
LWS
PMA / PMN Number
P920015
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT MEDICAL PRODUCTS: PRODUCT ID: D314DRG, IMPLANTED: (B)(6) 2013. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT EXPERIENCED SIXTY INAPPROPRIATE SHOCKS FROM THEIR IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD). OVERSENSING DUE TO LEAD FRACTURE NOISE WAS NOTED ON THE RIGHT VENTRICULAR (RV) LEAD. LOSS OF CAPTURE AND HIGH THRESHOLDS WERE ALSO INDICATED. THE LEAD WAS CAPPED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
638921 SPRINT QUATTRO SECURE DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS MPRI 694765

Patients

Seq Age Sex Outcome Treatment
1 00085 YR Hospitalization| L| R 5568 LEAD