FDA Adverse Event Malfunction Summary report: N

PULSE-GENERATOR, DUAL CHAMBER, IMPLANTABLE

MDR report key: 4160109 · Received October 10, 2014

Report

Report Number
2182208-2014-02863
Event Type
Malfunction
Date Received
October 10, 2014
Date of Event
May 2, 2014
Report Date
July 1, 2014
Manufacturer
MEDTRONIC, INC.
Product Code
LWP
PMA / PMN Number
P820003
Removal / Correction Number
Z-1661-2014
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THIS REPORT IS BASED SOLELY ON DEVICE ANALYSIS. NO INFORMATION TO SUGGEST A DEVICE RELATED ADVERSE EVENT OR PRODUCT PROBLEM WAS RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THIS DEVICE WAS INCLUDED IN THAT FIELD ACTION, BUT RETURNED PRODUCT TESTING FOUND THE DEVICE DID NOT PERFORM AS DESCRIBED IN THE FIELD ACTION. PRODUCT EVENT SUMMARY: ANALYSIS FOUND THAT THE HIGH RATE COVER WAS CONTAMINATED, THE MAIN CLEAR COVER WAS MISSING, THE LOWER CASE WAS BROKEN, THE BATTERY DRAWER END CAP WAS DENTED, THE BAIL COVER WAS CONTAMINATED, THE RING WAS BENT, AND THE INTERCONNECT FLEX WAS ROUTED INCORRECTLY. (B)(4).

Description of Event or Problem · 1

THE DEVICE WAS RETURNED TO THE MANUFACTURER, ANALYZED, AND TESTED OUT OF SPECIFICATION. NO PATIENT INVOLVEMENT OR COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.