FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 4160097 · Received September 17, 2014

Report

Report Number
2032227-2014-23915
Event Type
Malfunction
Date Received
September 17, 2014
Date of Event
August 18, 2014
Report Date
August 18, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP PASSED THE DISPLACEMENT, REWIND, BASIC OCCLUSION, OCCLUSION, PRIME, AND EXCESSIVE NO DELIVERY ALARM TEST. NO MOTOR ERROR ALARM WAS NOTED AND THE MOTOR PASSED THE MOTOR TEST. THE INSULIN PUMP WAS RECEIVED WITH MINOR SCRATCHES ON THE DISPLAY WINDOW AND A CRACKED RESERVOIR TUBE LIP.

Additional Manufacturer Narrative · 1

(B)(4). CURRENTLY, IT IS UNK WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OF CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFO WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. THE INSULIN PUMP INVOLVED IN THIS EVENT IS THE PARADIGM REAL-TIME VEO INSULIN INFUSION PUMP, WHICH IS NOT MARKETED IN THE UNITED STATES. HOWEVER, THE DEVICE IS SIMILAR TO THE PARADIGM REAL-TIME INSULIN INFUSION PUMP, WHICH IS MARKETED IN THE UNITED STATES.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULIN PUMP HAD ALARMED FOR A MOTOR ERROR DURING THE PRIME. THE BLOOD GLUCOSE READING WAS NOT GIVEN. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
574176 PARADIGM REAL-TIME INSULIN INFUSION PUMP OYC MEDTRONIC MINIMED MMT-554WWL

Patients

Seq Age Sex Outcome Treatment
1