FDA Adverse Event Injury Summary report: N

VIVA QUAD XT

MDR report key: 4160084 · Received October 10, 2014

Report

Report Number
9614453-2014-02400
Event Type
Injury
Date Received
October 10, 2014
Date of Event
June 12, 2014
Report Date
July 11, 2014
Manufacturer
IPG MFG SWITZERLAND
Product Code
NIK
PMA / PMN Number
P010031
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. THIS MODEL NUMBER IS NOT APPROVED FOR DISTRIBUTION IN THE UNITED STATES, HOWEVER, IT IS SIMILAR TO A DEVICE MARKETED IN THE U.S. THE EVENT IS BEING REPORTED DUE TO AN ALLEGED MALFUNCTION. (B)(4).

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE DEVICE WAS NOT RETURNED FOR ANALYSIS. HOWEVER, PERFORMANCE DATA COLLECTED FROM THE DEVICE WAS RECEIVED AND ANALYZED. ANALYSIS OF THE DEVICE MEMORY INDICATED THE ELECTRICAL RESET ALERT WAS TRIGGERED, AND A POWER ON RESET OCCURRED. THE ANALYST NOTED THAT AN ILLEGAL ADDRESS POWER ON RESET (POR) WAS RECORDED ON (B)(6) 2014.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE EXPERIENCED A POWER ON RESET (POR) AT NOTED VIA A MONITOR TRANSMISSION. THE DEVICE WAS REPROGRAMMED AND REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
638045 VIVA QUAD XT DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO NIK IPG MFG SWITZERLAND DTBA2QQ

Patients

Seq Age Sex Outcome Treatment
1 00059 YR Required Intervention