VIVA QUAD XT
Report
- Report Number
- 9614453-2014-02400
- Event Type
- Injury
- Date Received
- October 10, 2014
- Date of Event
- June 12, 2014
- Report Date
- July 11, 2014
- Manufacturer
- IPG MFG SWITZERLAND
- Product Code
- NIK
- PMA / PMN Number
- P010031
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- OTHER
Narratives
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. THIS MODEL NUMBER IS NOT APPROVED FOR DISTRIBUTION IN THE UNITED STATES, HOWEVER, IT IS SIMILAR TO A DEVICE MARKETED IN THE U.S. THE EVENT IS BEING REPORTED DUE TO AN ALLEGED MALFUNCTION. (B)(4).
PRODUCT EVENT SUMMARY: THE DEVICE WAS NOT RETURNED FOR ANALYSIS. HOWEVER, PERFORMANCE DATA COLLECTED FROM THE DEVICE WAS RECEIVED AND ANALYZED. ANALYSIS OF THE DEVICE MEMORY INDICATED THE ELECTRICAL RESET ALERT WAS TRIGGERED, AND A POWER ON RESET OCCURRED. THE ANALYST NOTED THAT AN ILLEGAL ADDRESS POWER ON RESET (POR) WAS RECORDED ON (B)(6) 2014.
IT WAS REPORTED THAT THE DEVICE EXPERIENCED A POWER ON RESET (POR) AT NOTED VIA A MONITOR TRANSMISSION. THE DEVICE WAS REPROGRAMMED AND REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 638045 | VIVA QUAD XT | DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO | NIK | IPG MFG SWITZERLAND | DTBA2QQ |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00059 YR | Required Intervention |