CEMENT, BONE, VERTEBROPLASTY
Report
- Report Number
- 1030489-2014-03993
- Event Type
- Malfunction
- Date Received
- October 9, 2014
- Report Date
- September 11, 2014
- Manufacturer
- MEDTRONIC SOFAMOR DANEK
- Product Code
- NDN
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- PHYSICIAN
Narratives
LITERATURE CITATION : CHONGYAN WANG; SHUNWU FAN; JUNHUI LIU; LETU SUYOU; ZHI SHAN; FENGDONG ZHAO. "BASIVERTEBRAL FORAMEN COULD BE C ONNECTED WITH INTRAVERTEBRAL CLEFT: A POTENTIAL RISK FACTOR OF CEMENT LEAKAGE IN PERCUTANEOUS KYPHOPLASTY". THE SPINE JOURNAL 14 (2014) 1551-1558. (B)(4).
IT WAS REPORTED THAT A RETROSPECTIVE STUDY WAS PERFORMED ON 164 PATIENTS TO ASSESS THE INCIDENCE RATE OF DIFFERENT TYPES OF CEMENT LEAKAGE IN PKP (PERCUTANEOUS KYPHOPLASTY PROCEDURES) WITH OR WITHOUT INTRAVERTEBRAL CLEFTS AND TO DETERMINE WHETHER THE BASIVERTEBRAL FORAMEN COULD BE CONNECTED TO THE INTRAVERTEBRAL CLEFT. BETWEEN JANUARY 2009 AND JANUARY 2012, 183 PATIENTS UNDERWENT PKP TO TREAT OVCFS. ALL PATIENTS REPORTED REFRACTORY LOW BACK PAIN AND TENDERNESS, AND FAILED TO RESPOND TO CONSERVATIVE TREATMENT FOR AT LEAST 2 MONTHS; 204 VERTEBRAE IN 164 PATIENTS (129 WOMEN AND 35 MEN) WITH A MEAN AGE OF 71 YEARS (RANGE, 56-91 YEARS) WERE SELECTED FOR THE STUDY. A TOTAL OF 204 VERTEBRAE IN 164 CONSECUTIVE PATIENTS WHO UNDERWENT PKP TO TREAT OSTEOPOROTIC VERTEBRAL COMPRESSION FRACTURES WERE CLASSIFIED INTO TWO PATTERNS BASED ON PREOPERATIVE RADIOGRAPHS, CT SCANS, AND/OR MR IMAGES OF THE TREATED LEVELS: CLEFT PATTERN (WITH AN INTRAVERTEBRAL CLEFT IN THE VERTEBRAL BODY) AND TRABECULAR PATTERN (WITHOUT INTRA VERTEBRAL CLEFTS). A TRABECULAR PATTERN COULD BE INFERRED IF THE OPAQUE CEMENT WAS INTERSPERSED THROUGHOUT THE TRABECULAR SPACE LIKE A SPONGE. THE OCCURRENCE AND TYPE OF CEMENT LEAKAGE WERE ASSESSED ON POSTOPERATIVE RADIOGRAPHS, CT SCANS, AND/OR MR IMAGES OF THE TREATED LEVELS. THE PATTERNS OF CEMENT LEAKAGE WERE CLASSIFIED INTO FIVE TYPES: TYPE A, THROUGH A CORTICAL DEFECT INTO PARASPINAL SOFT TISSUES; TYPE B, THROUGH THE BASIVERTEBRAL FORAMEN; TYPE C, VIA THE NEEDLE CHANNE L; TYPE D, THROUGH A CORTICAL DEFECT INTO THE DISC SPACE; AND TYPE E, VIA THE PARAVERTEBRAL VEIN. IT WAS REPORTED THAT ASYMPTOMATIC CEMENT EXTRAVASATION VIA THE NEEDLE CHANNEL (TYPE C) OCCURRED IN 12 VERTEBRAE. ACCORDING TO THE ARTICLE, "TYPE C LEAKS OCCURRED WHEN THE NEEDLE WAS WITHDRAWN TOO FAST, BEFORE THE CEMENT HAS HARDENED, SO CEMENT FILLED THE TRACT OF THE NEEDLE ON LATERAL RADIOGRAPHS". NO PATIENT COMPLICATIONS WERE REPORTED IN THE ARTICLE FOR THESE 12 TYPE C CEMENT LEAKS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 635587 | CEMENT, BONE, VERTEBROPLASTY | NDN | MEDTRONIC SOFAMOR DANEK | NA | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00071 YR | KYPHOPLASTY DEVICES |