PARADIGM REAL-TIME INSULIN INFUSION PUMP
Report
- Report Number
- 2032227-2014-35311
- Event Type
- Malfunction
- Date Received
- October 8, 2014
- Date of Event
- September 9, 2014
- Report Date
- September 9, 2014
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SP
- Reporter Occupation
- OTHER
Narratives
INSULIN PUMP HAD A BUTTON ERROR ALARM DUE TO CORRODED KEYPAD TRACE. INSULIN PUMP RECEIVED WITH MINOR SCRATCHES ON DISPLAY WINDOW, CRACKED CASE AT DISPLAY WINDOW CORNERS, CRACKED BELT CLIP SLOT AND CRACKED RESERVOIR TUBE LIP.
(B)(4).CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
IT WAS REPORTED THAT THE INSULIN PUMP ALARMED BUTTON ERROR WHILE TRYING TO DELIVER A BOLUS. IT WAS STATED THAT THE CUSTOMER REMOVED AND REINSERTED THE BATTERY AND THE ALARM PERSISTS. BLOOD GLUCOSE VALUE IS 225 MG/DL. NOTHING FURTHER REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 631805 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-754WWL |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |