FDA Adverse Event Malfunction Summary report: N

RIATA ACTIVE FIXATION

MDR report key: 4153681 · Received October 8, 2014

Report

Report Number
2938836-2014-16528
Event Type
Malfunction
Date Received
October 8, 2014
Date of Event
August 11, 2014
Manufacturer
ST. JUDE MEDICAL, INC., CRMD
Product Code
NVY
PMA / PMN Number
P950022
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. NO COMPLAINT RECEIVED WITH RETURN OF DEVICE. FAILURE (EVENT) OBSERVED DURING ANALYSIS.A PARTIAL LEAD WAS RETURNED IN FIVE SEGMENTS FOR ANALYSIS. INTERNAL INSULATION ABRASION WAS NOTED AT 40.5-41.3CM FROM THE DISTAL TIP. THE ETFE COATING WAS INTACT AT THIS LOCATION.

Description of Event or Problem · 1

THIS REPORT IS TO ADVISE OF AN EVENT OBSERVED DURING ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
631854 RIATA ACTIVE FIXATION DEFIBRILLATION LEAD NVY ST. JUDE MEDICAL, INC., CRMD 1581/65 NA

Patients

Seq Age Sex Outcome Treatment
1