FDA Adverse Event
Injury
Summary report: N
DURATA STS OPTIM ACTIVE FIXATION, DF-4 CONNECTOR
MDR report key: 4153608
·
Received October 8, 2014
Report
- Report Number
- 2938836-2014-16349
- Event Type
- Injury
- Date Received
- October 8, 2014
- Date of Event
- August 8, 2014
- Manufacturer
- ST. JUDE MEDICAL, INC., CRMD
- Product Code
- NVY
- PMA / PMN Number
- P950022
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WA, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. A PARTIAL LEAD WITH THE CONNECTOR PIN MEASURING 19.3CM WAS RETURNED FOR ANALYSIS. THE DAMAGE FOUND WAS SUSTAINED DURING THE SURGICAL PROCEDURE. THE PORTION OF THE LEAD THAT WAS RETURNED WAS OTHERWISE NORMAL. WITHOUT RETURN OF THE ENTIRE LEAD, A COMPLETE ANALYSIS COULD NOT BE PERFORMED.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE PATIENT PRESENTED IN CLINIC FOR FOLLOW UP. STORED EGMS REVEALED MULTIPLE EPISODES OF NOISE. THE NOISE COULD NOT BE REPRODUCED IN-CLINIC. THE LEAD WAS CAPPED AND REPLACED. PATIENT CONDITION WAS STABLE AFTER THE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 631627 | DURATA STS OPTIM ACTIVE FIXATION, DF-4 CONNECTOR | DEFIBRILLATION LEAD | NVY | ST. JUDE MEDICAL, INC., CRMD | 7120Q/58 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 70 YR | Required Intervention |