FDA Adverse Event Injury Summary report: N

DURATA STS OPTIM ACTIVE FIXATION, DF-4 CONNECTOR

MDR report key: 4153608 · Received October 8, 2014

Report

Report Number
2938836-2014-16349
Event Type
Injury
Date Received
October 8, 2014
Date of Event
August 8, 2014
Manufacturer
ST. JUDE MEDICAL, INC., CRMD
Product Code
NVY
PMA / PMN Number
P950022
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. A PARTIAL LEAD WITH THE CONNECTOR PIN MEASURING 19.3CM WAS RETURNED FOR ANALYSIS. THE DAMAGE FOUND WAS SUSTAINED DURING THE SURGICAL PROCEDURE. THE PORTION OF THE LEAD THAT WAS RETURNED WAS OTHERWISE NORMAL. WITHOUT RETURN OF THE ENTIRE LEAD, A COMPLETE ANALYSIS COULD NOT BE PERFORMED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT PRESENTED IN CLINIC FOR FOLLOW UP. STORED EGMS REVEALED MULTIPLE EPISODES OF NOISE. THE NOISE COULD NOT BE REPRODUCED IN-CLINIC. THE LEAD WAS CAPPED AND REPLACED. PATIENT CONDITION WAS STABLE AFTER THE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
631627 DURATA STS OPTIM ACTIVE FIXATION, DF-4 CONNECTOR DEFIBRILLATION LEAD NVY ST. JUDE MEDICAL, INC., CRMD 7120Q/58 NA

Patients

Seq Age Sex Outcome Treatment
1 70 YR Required Intervention