FDA Adverse Event
Malfunction
Summary report: N
RIATA ST ACTIVE FIXATION
MDR report key: 4153597
·
Received October 8, 2014
Report
- Report Number
- 2938836-2014-16516
- Event Type
- Malfunction
- Date Received
- October 8, 2014
- Date of Event
- September 10, 2014
- Manufacturer
- ST. JUDE MEDICAL, INC., CRMD
- Product Code
- NVY
- PMA / PMN Number
- P950022
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. NO COMPLAINT RECEIVED WITH RETURN OF DEVICE. FAILURE (EVENT) OBSERVED DURING ANALYSIS. EXTERNAL INSULATION ABRASION WAS NOTED AT 19.6-20.2CM AND 20.3-20.5CM FROM THE CONNECTOR PIN, CONSISTENT WITH LEAD FRICTION TO THE DEVICE CAN. THE ETFE COATING WAS INTACT AT THESE LOCATIONS. EXTERNAL INSULATION ABRASION WAS NOTED AT 18.6-19.3CM FROM THE CONNECTOR PIN, CONSISTENT WITH LEAD FRICTION TO THE DEVICE CAN. THE ETFE COATING WAS ABRADED AT THIS LOCATION.
Description of Event or Problem · 1
THIS REPORT IS TO ADVISE OF AN EVENT OBSERVED DURING ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 631737 | RIATA ST ACTIVE FIXATION | DEFIBRILLATION LEAD | NVY | ST. JUDE MEDICAL, INC., CRMD | 7010/65 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |