FDA Adverse Event
Malfunction
Summary report: N
UNIFY CRT-D
MDR report key: 4153595
·
Received October 8, 2014
Report
- Report Number
- 2938836-2014-16342
- Event Type
- Malfunction
- Date Received
- October 8, 2014
- Date of Event
- August 2, 2014
- Manufacturer
- ST. JUDE MEDICAL, INC., CRMD
- Product Code
- NIK
- PMA / PMN Number
- P030054
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- LA, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.
Description of Event or Problem · 1
IT WAS REPORTED THAT FAR R-WAVE OVERSENSING ON THE ATRIAL CHANNEL WAS OBSERVED, WHICH RESULTED IN SEVERAL INAPPROPRIATE MODE SWITCHES. PROGRAMMING CHANGES WERE MADE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 631898 | UNIFY CRT-D | IMPLANTABLE CARDIOVERTER DEFIBRILLATOR | NIK | ST. JUDE MEDICAL, INC., CRMD | CD3231-40 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 68 YR |