FDA Adverse Event Malfunction Summary report: N

QUADRA ASSURA CRT-D, DF-4 CONNECTOR

MDR report key: 4153553 · Received October 8, 2014

Report

Report Number
2938836-2014-16404
Event Type
Malfunction
Date Received
October 8, 2014
Date of Event
August 26, 2014
Manufacturer
ST. JUDE MEDICAL, INC., CRMD
Product Code
NIK
PMA / PMN Number
P910023
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.

Description of Event or Problem · 1

IT WAS REPORTED THAT NON-SUSTAINED LEAD NOISE DUE TO POST-PACED T-WAVE OVERSENSING WAS OBSERVED VIA REMOTE TRANSMISSION. PROGRAMMING CHANGES WERE MADE AND NO MORE ALERTS WERE SEEN. THE PATIENT WAS ASYMPTOMATIC AND DOING FINE AFTER THE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
632276 QUADRA ASSURA CRT-D, DF-4 CONNECTOR IMPLANTABLE CARDIOVERTER DEFIBRILLATOR NIK ST. JUDE MEDICAL, INC., CRMD CD3365-40Q NA

Patients

Seq Age Sex Outcome Treatment
1 69 YR