FDA Adverse Event
Malfunction
Summary report: N
QUADRA ASSURA DR CRT-D
MDR report key: 4153545
·
Received October 8, 2014
Report
- Report Number
- 2938836-2014-16506
- Event Type
- Malfunction
- Date Received
- October 8, 2014
- Date of Event
- August 21, 2014
- Manufacturer
- ST. JUDE MEDICAL, INC., CRMD
- Product Code
- NIK
- PMA / PMN Number
- P910023
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING A POST-OP CHECK UPON INTERROGATION, THE DEVICE SHOWED A DEVICE PARAMETER ERROR MESSAGE. THE DEVICE WAS SET BACK TO NOMINAL SETTINGS AND REPROGRAMMED. THERE WERE NO FURTHER COMPLICATIONS AND NO SYMPTOMS WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 632096 | QUADRA ASSURA DR CRT-D | IMPLANTABLE CARDIOVERTER DEFIBRILLATOR | NIK | ST. JUDE MEDICAL, INC., CRMD | CD3365-40C | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 66 YR |