FDA Adverse Event
Malfunction
Summary report: N
FORTIFY ASSURA DR
MDR report key: 4153505
·
Received October 8, 2014
Report
- Report Number
- 2938836-2014-16332
- Event Type
- Malfunction
- Date Received
- October 8, 2014
- Date of Event
- August 20, 2014
- Manufacturer
- ST. JUDE MEDICAL, INC., CRMD
- Product Code
- LWS
- PMA / PMN Number
- NA
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- NURSE
Narratives
Additional Manufacturer Narrative · 1
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.
Description of Event or Problem · 1
IT WAS REPORTED THAT UPON REVIEW OF AMS EPISDOES, POST-PACED T WAVE OVERSENSING WAS OBSERVED. PROGRAMMING CHANGES WERE MADE AND RESOLVED THE OVERSENSING. PATIENT IS DOING FINE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 630661 | FORTIFY ASSURA DR | IMPLANTABLE CARDIOVERTER DEFIBRILLATOR | LWS | ST. JUDE MEDICAL, INC., CRMD | CD2259-40 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | MEDT 5076 |