UNIFY ASSURA DR CRT-D

MDR report key: 4153501 · Received October 8, 2014

Report

Report Number
2938836-2014-16384
Event Type
Malfunction
Date Received
October 8, 2014
Date of Event
July 22, 2014
Manufacturer
ST. JUDE MEDICAL, INC., CRMD
Product Code
NIK
PMA / PMN Number
P030054
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.

Description of Event or Problem · 1

IT WAS REPORTED THAT MULTIPLE REMOTE TRANSMISSIONS REVEALED THAT LOW LEVEL HIGH FREQUENCY NOISE WAS BEING SENSED AS THE R-WAVE. PROGRAMMING CHANGES RESOLVED THE OVERSENSING. THE PATIENT WAS ASYMPTOMATIC.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
631167 UNIFY ASSURA DR CRT-D IMPLANTABLE CARDIOVERTER DEFIBRILLATOR NIK ST. JUDE MEDICAL, INC., CRMD CD3357-40C NA

Patients

Seq Age Sex Outcome Treatment
1 60 YR