FDA Adverse Event Malfunction Summary report: N

COULTER LH 750 HEMATOLOGY ANALYZER

MDR report key: 4153427 · Received October 8, 2014

Report

Report Number
1061932-2014-02492
Event Type
Malfunction
Date Received
October 8, 2014
Date of Event
September 11, 2014
Report Date
September 11, 2014
Manufacturer
BECKMAN COULTER
Product Code
GKZ
PMA / PMN Number
K061574
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE FSE FOUND DETACHED TUBING FROM PNEUMATIC VALVE (VL111) AND REATTACHED THE TUBING TO RESOLVE THE LEAK. THE FSE THEN PERFORMED VERIFICATION OF INSTRUMENT TO MEET THE SPECIFIED REQUIREMENTS PER ESTABLISHED PROCEDURES. BEC INTERNAL IDENTIFIER FOR THIS REPORT IS (B)(4).

Description of Event or Problem · 1

THE CUSTOMER REPORTED A CONTAINED FLUID LEAK CONSISTING OF DILUENT AND BLOOD MIXTURE FROM THE COULTER LH 750 HEMATOLOGY ANALYZER. THERE WERE NO REPORTS OF INJURIES, ERRONEOUS RESULTS, DIRECT PHYSICAL CONTACT WITH THE LEAK OR CHANGE TO PATIENT TREATMENT IN CONNECTION WITH THIS EVENT. THE CUSTOMER WAS WEARING PERSONAL PROTECTIVE EQUIPMENT CONSISTING OF GLOVES, GLASSES AND LABORATORY COAT WHEN THE LEAK OCCURRED. A BECKMAN COULTER (BEC) FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO THE CUSTOMER SITE TO EVALUATE THE INSTRUMENT.