FDA Adverse Event
Malfunction
Summary report: N
11CM ANGLE ATTACHMENT
MDR report key: 4153412
·
Received October 8, 2014
Report
- Report Number
- 1045834-2014-13321
- Event Type
- Malfunction
- Date Received
- October 8, 2014
- Report Date
- March 27, 2013
- Manufacturer
- DEPUY SYNTHES POWER TOOLS
- Product Code
- HBC
- PMA / PMN Number
- PK011444
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
SYNTHES IS SUBMITTING THIS REPORT AS A RESULT OF REMEDIATION ACTIVITIES RELATED TO SYNTHES OUS SERVICE AND REPAIR FILES LEGACY REVIEW/REMEDIATION PROTOCOL-COMPLAINT HANDLING AND MDR REPORTING. DEVICE LISTED IN THIS REPORT IS USED FOR TREATMENT, NOT DIAGNOSIS. ANY ADDITIONAL INFORMATION RECEIVED REGARDING THIS EVENT AFTER FILING THIS REPORT SHALL BE FILED ON A SUPPLEMENTAL MDR. DURING PRE-REPAIR ASSESSMENT PERFORMED BY TECHNICIAN, INVESTIGATION REVEALED THE DEVICE HAS A DEFECTIVE BEARING AND GUIDE TUBE. THE DEVICE WAS REPAIRED AND RETURNED TO CUSTOMER ON (B)(4) 2013. (B)(4).
Description of Event or Problem · 1
CUSTOMER REPORTED THE TOP OF DEVICE OVERHEATS. THIS IS REPORT 1 OF 1 FOR (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 631320 | 11CM ANGLE ATTACHMENT | MOTOR, DRILL, ELECTRIC | HBC | DEPUY SYNTHES POWER TOOLS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |