FDA Adverse Event Malfunction Summary report: N

11CM ANGLE ATTACHMENT

MDR report key: 4153393 · Received October 8, 2014

Report

Report Number
1045834-2014-13153
Event Type
Malfunction
Date Received
October 8, 2014
Date of Event
September 25, 2014
Report Date
September 25, 2014
Manufacturer
DEPUY SYNTHES POWER TOOLS
Product Code
HBC
PMA / PMN Number
PK011444
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE ACTUAL DEVICE WAS RETURNED FOR EVALUATION. RELIABILITY ENGINEERING EVALUATED THE DEVICE AND FOUND THAT THE UNIT FAILED CUTTER INSERTION. THEREFORE, THE REPORTED CONDITION WAS CONFIRMED. IT WAS DETERMINED THAT THE BEARINGS WERE WORN AND NO LONGER IN AXIAL ALIGNMENT. THE ASSIGNABLE ROOT CAUSE WAS DETERMINED TO BE DUE TO NORMAL WEAR AND SERVICING OVER TIME. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING PRE-SURGERY, IT WAS OBSERVED THAT THE DRILL BIT WOULD NOT TURN IN ATTACHMENT DEVICE. THERE WERE NO DELAYS TO THE PLANNED SURGICAL PROCEDURE AS AN IDENTICAL SPARE DEVICE WAS AVAILABLE FOR USE. THERE WAS NO PATIENT INVOLVEMENT REPORTED. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.