FDA Adverse Event Malfunction Summary report: N

HOMECHOICE AUTOMATED PD SET WITH CASSETTE

MDR report key: 4153391 · Received October 8, 2014

Report

Report Number
1416980-2014-35071
Event Type
Malfunction
Date Received
October 8, 2014
Report Date
September 16, 2014
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FKX
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RECEIVED FOR EVALUATION. VISUAL INSPECTION WAS PERFORMED AND NOTED CRACKS ON THE SIDE OF THE CASSETTE WITHOUT ANY SIGN OF PHYSICAL DAMAGE. PRESSURE TESTING WAS PERFORMED AND A LEAK WAS VERIFIED AT THE CRACK ON THE SIDE OF THE CASSETTE. SAMPLE ANALYSIS VERIFIED THE REPORTED EVENT. A REVIEW OF ALL BATCH RECORD DOCUMENTS WAS PERFORMED FOR ASSOCIATED LOT NUMBER S14D07068 WITH NO ISSUES NOTED DURING THE MANUFACTURING PROCESS. THERE WERE NO DEVIATIONS FROM STANDARD PROCEDURE AND NO EXCEPTIONS RELATED TO THE REPORTED CONDITION WERE NOTED. UPON CONCLUSION OF THE INVESTIGATION, THE CAUSE OF THE EVENT WAS A CRACKED CASSETTE. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS NOT RETURNED; THEREFORE, A DEVICE ANALYSIS COULD NOT BE COMPLETED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A HOME PATIENT (HP) EXPERIENCED A LEAK FROM THE CASSETTE ON THE HOMECHOICE DEVICE, DURING PRIMING. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION INDICATED AT THE TIME OF THE REPORT. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
631306 HOMECHOICE AUTOMATED PD SET WITH CASSETTE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - SINGAPORE S14D07068

Patients

Seq Age Sex Outcome Treatment
1 HOMECHOICE