FDA Adverse Event Malfunction Summary report: N

COLIBRI

MDR report key: 4153372 · Received October 8, 2014

Report

Report Number
8030965-2014-00518
Event Type
Malfunction
Date Received
October 8, 2014
Report Date
May 6, 2013
Manufacturer
SYNTHES GMBH
Product Code
MOQ
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

SYNTHES IS SUBMITTING THIS REPORT AS A RESULT OF REMEDIATION ACTIVITIES RELATED TO SYNTHES OUS SERVICE AND REPAIR FILES LEGACY REVIEW/REMEDIATION PROTOCOL-COMPLAINT HANDLING AND MDR REPORTING. DEVICE LISTED IN THIS REPORT IS USED FOR TREATMENT, NOT DIAGNOSIS. ANY ADDITIONAL INFORMATION RECEIVED REGARDING THIS EVENT AFTER FILING THIS REPORT SHALL BE FILED ON A SUPPLEMENTAL MDR. DURING PRE-REPAIR ASSESSMENT PERFORMED BY TECHNICIAN, INVESTIGATION REVEALED THE MOTOR WAS BLOCKED, ROUGH RUNNING AND SEIZED. THE DEVICE WAS REPAIRED AND RETURNED TO CUSTOMER ON (B)(4) 2013. (B)(6). PLACEHOLDER.

Description of Event or Problem · 1

CUSTOMER REPORTED THE MOTOR MAKES LOUD NOISE. THIS IS REPORT 1 OF 1 FOR (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
630549 COLIBRI MOQ SYNTHES GMBH 9408

Patients

Seq Age Sex Outcome Treatment
1