FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 4153214 · Received October 8, 2014

Report

Report Number
2531779-2014-28666
Event Type
Malfunction
Date Received
October 8, 2014
Report Date
September 27, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 09/27/2014 WITH THE FOLLOWING FINDINGS: THE BATTERY COMPARTMENT WAS OBSERVED TO BE CRACKED BETWEEN THE BUMPER PAD AND THE COVER. THE DISPLAY SCREEN HAD A PINKISH CONTRAST. THE RETURNED BATTERY CAP WAS ABLE TO FULLY ATTACH TO THE PUMP AND WAS USED TO COMPLETE ALL TESTING. THE BLACK BOX INDICATED A TIME AND DATE RESET EVENT HAD OCCURRED WITHIN 3 MINUTES OF POWERING THE PUMP OFF. THE PUMP WAS EXERCISED FOR 24 HOURS; THE BATTERY WAS THEN REMOVED FOR 6 HOURS AND THEN REINSERTED. THE TIME AND DATE WAS OBSERVED TO HAVE RESET TO 12:00AM (B)(4) 2007, WHICH INDICATED THAT AN INTERNAL CLOCK BATTERY MALFUNCTION HAD OCCURRED, AND WAS CONFIRMED BY A VISIBLE INSPECTION WHEN THE COVER WAS REMOVED. (B)(6).

Description of Event or Problem · 1

THE PUMP WAS RETURNED FOR INVESTIGATION. INVESTIGATION REVEALED A CRACKED BATTERY COMPARTMENT AND A PINK DISPLAY SCREEN. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON 09/27/2014.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
630753 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 33 YR