FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 4153021 · Received October 8, 2014

Report

Report Number
2531779-2014-28662
Event Type
Malfunction
Date Received
October 8, 2014
Report Date
September 27, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED, AND IT WAS EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2014 WITH THE FOLLOWING FINDINGS: THE OCCURRENCE OF A CS-064 CALL SERVICE ALARM, WHICH CAN BE INDICATIVE OF THE TYPE OF ISSUE THAT WAS REPORTED, WAS OBSERVED IN THE BLACK BOX AND HISTORY DATA. THE PUMP WAS EXERCISED FOR 24 HOURS, DURING WHICH NO CS-064 CALL SERVICE ALARMS WERE OBSERVED: THE REPORTED ISSUE WAS NOT DUPLICATED. THE PUMP CASE WAS REMOVED, AND NO EVIDENCE OF INTERNAL DAMAGE OR DEFECT WAS FOUND. DURING ANALYSIS, THE PUMP OPERATED ACCORDING TO THE SPECIFICATIONS. A BATTERY CAP WAS NOT RETURNED WITH THE PUMP; A TEST BATTERY CAP WAS USED DURING THE ANALYSIS.

Description of Event or Problem · 1

ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A CALL SERVICE ALARM (CS 064) ISSUE. IT WAS REPORTED THAT A CS-064 CALL SERVICE ALARM OCCURRED WHILE THE USER WAS PERFORMING A RANDOM FUNCTION. REPORTEDLY, THE USER HAD NOT RECEIVED MORE THAN ONE CS-064 CALL SERVICE ALARM OR OCCLUSION FOR THE SITE THEY WERE USING AT THAT TIME. THIS COMPLAINT IS BEING REPORTED BECAUSE THE ALLEGED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
631775 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 35 YR