FDA Adverse Event Malfunction Summary report: N

ABSORB BIORESORBABLE VASCULAR SCAFFOLD SYSTEM

MDR report key: 4153018 · Received October 8, 2014

Report

Report Number
2024168-2014-06520
Event Type
Malfunction
Date Received
October 8, 2014
Date of Event
September 15, 2014
Report Date
September 15, 2014
Manufacturer
AV-TEMECULA-CT
Product Code
NIQ
PMA / PMN Number
P070015
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE IS EXPECTED TO BE RETURNED FOR EVALUATION. IT HAS NOT YET BEEN RECEIVED. A FOLLOW UP REPORT WILL BE SUBMITTED WITH ALL RELEVANT INFORMATION. THOUGH THE DEVICE IS NOT APPROVED FOR SALE IN THE U.S., IT USES A DELIVERY SYSTEM WHICH IS SIMILAR TO A DEVICE SOLD IN THE U.S.

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE DEVICE WAS RETURNED FOR EVALUATION. THE TIP SEPARATION WAS NOT CONFIRMED; HOWEVER, THE BALLOON/SHAFT WAS SEPARATED. THE REPORTED DIFFICULTY REMOVING THE PROTECTIVE SHEATHS COULD NOT BE REPLICATED IN A TESTING ENVIRONMENT DUE TO THE CONDITION OF THE RETURNED DEVICE. BASED ON VISUAL AND DIMENSIONAL ANALYSIS OF THE RETURNED DEVICE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES THAT WOULD HAVE CONTRIBUTED TO THIS COMPLAINT. A QUERY/REVIEW OF THE COMPLAINT HISTORY OF THE REPORTED LOT DID NOT INDICATE A MANUFACTURING ISSUE. BASED ON THE INFORMATION REVIEWED, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE YELLOW PROTECTIVE SHEATH COULD NOT BE PROPERLY REMOVED FROM THE DELIVERY SYSTEM. DUE TO THE HEAVY RESISTANCE, THE TIP BECAME SEPARATED. THE DEVICE WAS NOT USED. THERE WAS NO PATIENT INVOLVEMENT. A NEW DEVICE WAS USED TO SUCCESSFULLY COMPLETE THE PROCEDURE. THERE WAS NO CLINICALLY SIGNIFICANT DELAY IN PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
632018 ABSORB BIORESORBABLE VASCULAR SCAFFOLD SYSTEM DRUG ELUTING CORONARY STENT SYSTEM NIQ AV-TEMECULA-CT 4020661

Patients

Seq Age Sex Outcome Treatment
1