FDA Adverse Event Malfunction Summary report: N

HOMECHOICE AUTOMATED PD SET WITH CASSETTE

MDR report key: 4153010 · Received October 8, 2014

Report

Report Number
1416980-2014-35043
Event Type
Malfunction
Date Received
October 8, 2014
Date of Event
September 15, 2014
Report Date
September 15, 2014
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
PMA / PMN Number
K102936
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
ND, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

COMPLAINT NO: CMPLNT-(B)(4). THE DEVICE WAS RECEIVED FOR EVALUATION. VISUAL INSPECTION OF THE DEVICE FOUND NO ISSUES. LEAK TESTING, CLAMP FUNCTION TESTING, CLEAR PASSAGE TESTING, AND A SIMULATED THERAPY WERE PERFORMED WITH THE DEVICE AND FOUND NO ISSUES. A REVIEW OF ALL BATCH RECORD DOCUMENTS WAS PERFORMED FOR LOT NUMBER H14F30016 WITH NO ISSUES NOTED DURING THE MANUFACTURING PROCESS. THERE WERE NO DEVIATIONS FROM STANDARD PROCEDURE AND NO EXCEPTIONS RELATED TO THE REPORTED CONDITION WERE NOTED. UPON CONCLUSION OF THE INVESTIGATION, THE CAUSE OF THE REPORTED EVENT COULD NOT BE DETERMINED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A HOME PATIENT (HP) EXPERIENCED A LEAK IN THE CASSETTE ON THE HOMECHOICE DEVICE, DURING USE. THE HP WAS NOT CONNECTED AT THE TIME OF THE EVENT. THE LEAK OF THE CASSETTE WAS FURTHER DESCRIBED AS ONE OF THE UNUSED SUPPLY LINES LEAKING FLUID EVEN THOUGH THE CLAMP WAS CLOSED ON THE LINE. DURING TROUBLESHOOTING, THERE WAS NOTHING UNUSUAL NOTED THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE TECHNICAL SERVICE REPRESENTATIVE ADVISED THE HP TO BEGIN THERAPY AGAIN WITH NEW SUPPLIES. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION INDICATED AT THE TIME OF THE REPORT. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
632262 HOMECHOICE AUTOMATED PD SET WITH CASSETTE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - MOUNTAIN HOME H14F30016

Patients

Seq Age Sex Outcome Treatment
1 85 YR HOMECHOICE