ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2014-28642
- Event Type
- Malfunction
- Date Received
- October 8, 2014
- Report Date
- September 26, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MD, US
- Reporter Occupation
- PATIENT
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 11/22/2014 WITH THE FOLLOWING FINDINGS: THE POWER COMPLAINT COULD NOT BE DUPLICATED. A REVIEW OF THE PUMP¿S BLACK BOX DATA REVEALED UNEXPLAINED PUMP REBOOTS. THE BATTERY COMPARTMENT WAS FOUND TO BE CRACKED. THERE WAS NO VISIBLE DAMAGE TO THE RETURNED BATTERY CAP, AND THE CAP WAS ABLE TO FULLY SECURE ONTO THE PUMP. THE PUMP WAS EXERCISED FOR 24 HOURS WITH NO POWER INTERRUPTIONS OR DUPLICATED ALARMS. THE LEAK TEST FAILED DUE TO A BATTERY COMPARTMENT LEAK. THE PUMP WAS OPENED AND NO EVIDENCE OF MOISTURE OR LOOSE COMPONENTS WAS FOUND TO THE INTERNAL COMPONENTS OF THE PUMP. UNRELATED TO THE INITIAL COMPLAINT, THERE WAS EVIDENCE OF CORROSION IN THE BATTERY COMPARTMENT. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A POWER (DAMAGE) ISSUE. THE REPORTER ALLEGED THAT THE BATTERY COMPARTMENT WAS CRACKED. IT WAS REPORTED THAT THE TOP OF THE BATTERY CAP WAS WORN, AND THE CAP HAD NOT BEEN REPLACED IN FOUR TO SIX MONTHS. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 631486 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 42 YR |