SPECTRA
Report
- Report Number
- 3006630150-2014-02113
- Event Type
- Malfunction
- Date Received
- October 8, 2014
- Date of Event
- June 12, 2014
- Report Date
- June 12, 2014
- Manufacturer
- BOSTON SCIENTIFIC NEUROMODULATION
- Product Code
- LGW
- PMA / PMN Number
- 030017
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AS
- Reporter Occupation
- PHYSICIAN
Narratives
ADDITIONAL SUSPECT MEDICAL DEVICE COMPONENTS INVOLVED IN THE EVENT: MODEL #: SC-4316, LOT #: 16377029, DESCRIPTION: NEXT GENERATION ANCHOR, KIT-STERILE; MODEL #: SC-2218-70, SERIAL #: (B)(4), DESCRIPTION: LINEAR ST LEAD, 70CM. VISUAL AND X-RAY INSPECTIONS REVEALED: LEAD SC-2218-70 (SN (B)(4)); ALL CABLES WERE FRACTURED AT THE CLICK ANCHOR SITE. LEAD SC-2218-70(SN (B)(4)); FOUR CABLES WERE FRACTURED AT THE CLIK ANCHOR SITE. LEADS SC-2218-70 (SN (B)(4)); REVEALED NO ANOMALIES. THE LEADS PASSED ELECTRICAL, PHOTOGRAPHIC AND VISUAL TESTING. - ALL FOUR CLIK ANCHORS HAD DAMAGED EYELETS. EACH CLIK ANCHOR IS MISSING A PORTION OF THE DAMAGED EYELETS. THE SMALL PIECES OF SILICONE WERE NOT RETURNED TO BSN. NO PIECES OF THE CLIK ANCHOR WERE LEFT INSIDE THE PATIENT.
(B)(4). VISUAL AND X-RAY INSPECTIONS REVEALED: LEAD SC-2218-70 (SN 430163); ALL CABLES WERE FRACTURED AT THE CLICK ANCHOR SITE. LEAD SC-2218-70 (SN (B)(4)); FOUR CABLES WERE FRACTURED AT THE CLIK ANCHOR SITE. LEADS SC-2218-70 ((B)(4)); REVEALED NO ANOMALIES. THE LEADS PASSED ELECTRICAL, PHOTOGRAPHIC AND VISUAL TESTING. ALL FOUR CLIK ANCHORS HAD DAMAGED EYELETS. EACH CLIK ANCHOR IS MISSING A PORTION OF THE DAMAGED EYELETS. THE SMALL PIECES OF SILICONE WERE NOT RETURNED. ADDITIONAL SUSPECT MEDICAL DEVICE COMPONENTS INVOLVED IN THE EVENT: MODEL #: SC-4316, LOT #: 16377029, DESCRIPTION: NEXT GENERATION ANCHOR, KIT-STERILE.
A REPORT WAS RECEIVED THAT FOLLOWING A LEAD REVISION PROCEDURE DUE TO LEAD MIGRATION AND HIGH IMPEDANCES, THE LEADS WERE RETURNED, ANALYZED AND FOUND TO BE BROKEN. IN ADDITION, THE CLIK ANCHORS HAD DAMAGED EYELETS.
A REPORT WAS RECEIVED THAT FOLLOWING A LEAD REVISION PROCEDURE DUE TO LEAD MIGRATION AND HIGH IMPEDANCES, THE LEADS WERE RETURNED, ANALYZED AND FOUND TO BE BROKEN. IN ADDITION, THE CLIK ANCHORS HAD DAMAGED EYELETS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 632220 | SPECTRA | SPINAL CORD STIMULATOR | LGW | BOSTON SCIENTIFIC NEUROMODULATION | SC-2218-70 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |