FDA Adverse Event Malfunction Summary report: N

SPECTRA

MDR report key: 4152966 · Received October 8, 2014

Report

Report Number
3006630150-2014-02113
Event Type
Malfunction
Date Received
October 8, 2014
Date of Event
June 12, 2014
Report Date
June 12, 2014
Manufacturer
BOSTON SCIENTIFIC NEUROMODULATION
Product Code
LGW
PMA / PMN Number
030017
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

ADDITIONAL SUSPECT MEDICAL DEVICE COMPONENTS INVOLVED IN THE EVENT: MODEL #: SC-4316, LOT #: 16377029, DESCRIPTION: NEXT GENERATION ANCHOR, KIT-STERILE; MODEL #: SC-2218-70, SERIAL #: (B)(4), DESCRIPTION: LINEAR ST LEAD, 70CM. VISUAL AND X-RAY INSPECTIONS REVEALED: LEAD SC-2218-70 (SN (B)(4)); ALL CABLES WERE FRACTURED AT THE CLICK ANCHOR SITE. LEAD SC-2218-70(SN (B)(4)); FOUR CABLES WERE FRACTURED AT THE CLIK ANCHOR SITE. LEADS SC-2218-70 (SN (B)(4)); REVEALED NO ANOMALIES. THE LEADS PASSED ELECTRICAL, PHOTOGRAPHIC AND VISUAL TESTING. - ALL FOUR CLIK ANCHORS HAD DAMAGED EYELETS. EACH CLIK ANCHOR IS MISSING A PORTION OF THE DAMAGED EYELETS. THE SMALL PIECES OF SILICONE WERE NOT RETURNED TO BSN. NO PIECES OF THE CLIK ANCHOR WERE LEFT INSIDE THE PATIENT.

Additional Manufacturer Narrative · 1

(B)(4). VISUAL AND X-RAY INSPECTIONS REVEALED: LEAD SC-2218-70 (SN 430163); ALL CABLES WERE FRACTURED AT THE CLICK ANCHOR SITE. LEAD SC-2218-70 (SN (B)(4)); FOUR CABLES WERE FRACTURED AT THE CLIK ANCHOR SITE. LEADS SC-2218-70 ((B)(4)); REVEALED NO ANOMALIES. THE LEADS PASSED ELECTRICAL, PHOTOGRAPHIC AND VISUAL TESTING. ALL FOUR CLIK ANCHORS HAD DAMAGED EYELETS. EACH CLIK ANCHOR IS MISSING A PORTION OF THE DAMAGED EYELETS. THE SMALL PIECES OF SILICONE WERE NOT RETURNED. ADDITIONAL SUSPECT MEDICAL DEVICE COMPONENTS INVOLVED IN THE EVENT: MODEL #: SC-4316, LOT #: 16377029, DESCRIPTION: NEXT GENERATION ANCHOR, KIT-STERILE.

Description of Event or Problem · 1

A REPORT WAS RECEIVED THAT FOLLOWING A LEAD REVISION PROCEDURE DUE TO LEAD MIGRATION AND HIGH IMPEDANCES, THE LEADS WERE RETURNED, ANALYZED AND FOUND TO BE BROKEN. IN ADDITION, THE CLIK ANCHORS HAD DAMAGED EYELETS.

Description of Event or Problem · 1

A REPORT WAS RECEIVED THAT FOLLOWING A LEAD REVISION PROCEDURE DUE TO LEAD MIGRATION AND HIGH IMPEDANCES, THE LEADS WERE RETURNED, ANALYZED AND FOUND TO BE BROKEN. IN ADDITION, THE CLIK ANCHORS HAD DAMAGED EYELETS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
632220 SPECTRA SPINAL CORD STIMULATOR LGW BOSTON SCIENTIFIC NEUROMODULATION SC-2218-70 NA

Patients

Seq Age Sex Outcome Treatment
1