FDA Adverse Event Malfunction Summary report: N

SYSTEM, PERITONEAL, AUTOMATIC DELIVERY

MDR report key: 4152808 · Received October 8, 2014

Report

Report Number
1416980-2014-34998
Event Type
Malfunction
Date Received
October 8, 2014
Date of Event
September 13, 2014
Report Date
September 13, 2014
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE CAUSE OF THE REPORTED EVENT WAS THAT THE SUPPLY BAG DISCONNECTED WITHOUT FALLING. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

(B)(4). THE SAMPLE WAS NOT RETURNED AND THE LOT NUMBER IS UNKNOWN; THEREFORE, A DEVICE EVALUATION COULD NOT BE PERFORMED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A HOME PATIENT (HP) HAD A SYSTEM ERROR 2240 (AIR IN TUBING) ALARM ON THE HOMECHOICE (HC) DURING INITIAL DRAIN. THE PATIENT WAS CONNECTED AT THE TIME OF THE ALARM. THIS OCCURRED DURING DWELL CYCLE TWO OF FOUR OF PERITONEAL DIALYSIS (PD) THERAPY. DURING THERAPY THE PATIENT¿S SUPPLY BAG DISCONNECTED FROM THE SUPPLY LINE WITHOUT FALLING, CAUSING THE ALARM. THE PATIENT COMPLETED THERAPY WITH A MANUAL EXCHANGE. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
630600 SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1 55 YR HOMECHOICE