GREENLIGHT MOXY FIBER OPTIC
Report
- Report Number
- 2937094-2014-00913
- Event Type
- Malfunction
- Date Received
- October 8, 2014
- Date of Event
- October 2, 2014
- Report Date
- October 2, 2014
- Manufacturer
- AMERICAN MEDICAL SYSTEMS, INC. (SAN JOSE)
- Product Code
- GEX
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SP
- Reporter Occupation
- OTHER
Narratives
(B)(4) (NO CODE AVAILABLE) REFERS TO FORWARD FIRING OF THE SIDE FIRING SURGICAL FIBER; A CODE REQUEST HAS BEEN SUBMITTED.
PRODUCT EVALUATION: FAILURE ANALYSIS FOR FIBER (B)(4): THE FIBER/GLASS CAP SHOWS A CIRCUMFERENTIAL FRACTURE DISTAL TO FIBER/CAP FUSION ZONE AT THE BEVEL EDGE; THE FIBER PROXIMAL TO FRACTURE CAN ROTATE INDEPENDENTLY OF METAL CAP; THE GLASS CAP EXHIBITS MILD DEVITRIFICATION AT THE OUTPUT WINDOW; THE METAL CAP EXHIBITS MILD CHAR; THE METAL CAP EXHIBITS SCRATCH MARKS. BASED ON THE DEVICE ANALYSIS, THE POTENTIAL FOR FORWARD FIRING MAY EXIST. PROBABLE ROOT CAUSE: BASED ON THE PRODUCT ANALYSIS RESULTS, THE PROBABLE ROOT CAUSE OF THE FAILURE IS: HEAT ACCUMULATION. CAP WEAR WAS ACCELERATED DUE TO ANATOMICAL/PROCEDURAL FACTORS (TISSUE CONTACT AND TECHNIQUE) ENCOUNTERED DURING THE PROCEDURE WHICH WOULD LIMIT THE PERFORMANCE OF THE FIBER.
IT WAS REPORTED THAT DURING A PROSTATE PROCEDURE, FORWARD FIRING WAS OBSERVED @ 18:36 MINUTES OF USE. 83,516 JOULES. THERE WERE "NO DAMAGES TO THE PATIENT" REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 630527 | GREENLIGHT MOXY FIBER OPTIC | POWERED LASER SURGICAL INSTRUMENT | GEX | AMERICAN MEDICAL SYSTEMS, INC. (SAN JOSE) | 0010-2400 | 407A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |