GEMSTAR PM THERAPY S
Report
- Report Number
- 9615050-2014-05505
- Event Type
- Malfunction
- Date Received
- October 6, 2014
- Date of Event
- September 16, 2014
- Report Date
- September 16, 2014
- Manufacturer
- HOSPIRA COSTA RICA LTD.
- Product Code
- FRN
- PMA / PMN Number
- K083019
- Removal / Correction Number
- Z-1170-2013
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IA, US
- Reporter Occupation
- OTHER
Narratives
TESTING AND INVESTIGATION FOUND WHILE RUNNING THE CUSTOMER'S LAST PROTOCOL THE DEVICE WOULD POWER DOWN WHEN IT IS KNOCKED ONTO A BENCH WHILE RUNNING ON DC POWER. IT WAS OBSERVED THAT WHEN THE POWER LOSS OCCURS, THE DEVICE WOULD ALARM A CHIRP TONE. THIS WAS TESTED THREE TIMES WITH THE SAME RESULTS. VISUAL INSPECTION FOUND THAT THE BATTERY CONTACT J104 OF THE MIDDLE PWA WAS LOOSE AND THERE WAS EVIDENCE OF CORROSION. THIS MAY CAUSE INTERMITTENT CONTACT BETWEEN THE AA BATTERIES AND THE POSITIVE BATTERY CONTACT WHICH WILL RESULT IN INTERMITTENT POWER LOOSES. A VISUAL INSPECTION ALSO FOUND CORROSION ON THE CASE ASSEMBLY. THE PROBABLE CAUSE OF THE POWER LOSS MAY BE DUE TO A LOOSE BATTERY CONTACT J104 ON THE MIDDLE PWA. THERE WAS ALSO CONTAMINATION ON THE POSITIVE BATTERY CONTACT J104. THE PROBABLE CAUSE WAS DUE TO LEAKAGE FROM THE DISPOSABLE BATTERIES. THIS REPORT REPRESENTS ALL THE INFORMATION KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.
THE CUSTOMER CONTACT REPORTED THAT DURING PREVENTIVE MAINTENANCE TESTING AT THE USER FACILITY, CORROSION DUE TO BATTERY LEAKAGE WAS NOTED IN THE BATTERY COMPARTMENT OF THE DEVICE. FURTHER TESTING FOUND THAT THE DEVICE POWERED OFF BY ITSELF WITHOUT SOUNDING AN AUDIBLE ALARM TONE WHEN IT WAS BUMPED. THERE WERE NO REPORTS OF ANY ADVERSE PATIENT EVENTS OR DELAYS IN CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 622200 | GEMSTAR PM THERAPY S | 80FRN | FRN | HOSPIRA COSTA RICA LTD. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |