FDA Adverse Event Malfunction Summary report: N

GEMSTAR PUMP PAIN MA

MDR report key: 4152577 · Received October 6, 2014

Report

Report Number
9615050-2014-05501
Event Type
Malfunction
Date Received
October 6, 2014
Date of Event
August 28, 2014
Report Date
September 3, 2014
Manufacturer
HOSPIRA COSTA RICA LTD.
Product Code
FRN
PMA / PMN Number
K083019
Removal / Correction Number
Z-1170-2013
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DURING TESTING AT THE SERVICE CENTER CORROSION WAS NOTED IN THE BATTERY COMPARTMENT OF THE DEVICE. THIS WAS DUE TO BATTERY LEAKAGE FROM THE DISPOSABLE BATTERIES. THE DEVICE HAS BEEN IDENTIFIED AS PART OF A PRODUCT RECALL. THIS REPORT REPRESENTS ALL THE INFORMATION KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.

Description of Event or Problem · 1

THE DEVICE WAS RETURNED TO THE SERVICE CENTER WITH A REPORT FROM THE CUSTOMER CONTACT THAT THE DEVICE ALARMED WITH AN UNSPECIFIED ALARM CODE. NO ADDITIONAL INFORMATION WAS PROVIDED. THIS IS NOT A REPORTABLE MALFUNCTION. HOWEVER, DURING VERIFICATION TESTING AT THE SERVICE CENTER, LEAKAGE FROM THE DISPOSABLE BATTERIES WAS NOTED IN THE BATTERY COMPARTMENT OF THE DEVICE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
622204 GEMSTAR PUMP PAIN MA 80FRN FRN HOSPIRA COSTA RICA LTD. NA NA

Patients

Seq Age Sex Outcome Treatment
1 NA