FDA Adverse Event Malfunction Summary report: N

SYMBIQ SINGLE CHANNE

MDR report key: 4152573 · Received October 6, 2014

Report

Report Number
9615050-2014-05494
Event Type
Malfunction
Date Received
October 6, 2014
Date of Event
January 1, 2014
Report Date
September 18, 2014
Manufacturer
HOSPIRA COSTA RICA LTD.
Product Code
FRN
PMA / PMN Number
K110901
Removal / Correction Number
Z-0069-2013
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
HI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE FIELD SERVICE ENGINEER PERFORMED TESTING AND INVESTIGATION AT THE USER FACILITY, THE DEVICE TOUCHSCREEN DID NOT RESPOND TO ACCESS THE BIOMEDICAL MODE AND DID NOT PASS THE TOUCHSCREEN CALIBRATION. THE PROBABLE CAUSE WAS DUE TO A BROKEN DEVICE TOUCHSCREEN. AS INDICATED, THIS DEVICE HAS BEEN IDENTIFIED AS PART OF A PRODUCT RECALL. THIS REPORT REPRESENTS ALL THE INFORMATION KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.

Description of Event or Problem · 1

THE CUSTOMER CONTACT REPORTED THE DEVICE TOUCHSCREEN MODE BUTTON DID NOT RESPOND. THERE WERE NO REPORTS OF ANY ADVERSE PATIENT EVENTS OR DELAYS IN CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. DURING TESTING AT THE USER FACILITY, THE DEVICE TOUCHSCREEN DID NOT RESPOND TO ACCESS THE BIOMEDICAL MODE. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
622146 SYMBIQ SINGLE CHANNE 80FRN FRN HOSPIRA COSTA RICA LTD. NA NA

Patients

Seq Age Sex Outcome Treatment
1 NA