FDA Adverse Event Malfunction Summary report: N

SYMBIQ DCHANNEL 3.01

MDR report key: 4152440 · Received October 3, 2014

Report

Report Number
9615050-2014-05476
Event Type
Malfunction
Date Received
October 3, 2014
Date of Event
January 1, 2014
Report Date
September 18, 2014
Manufacturer
HOSPIRA COSTA RICA LTD.
Product Code
FRN
PMA / PMN Number
K041550
Removal / Correction Number
Z0070-2013
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A FIELD SERVICE ENGINEER PERFORMED TESTING AND INVESTIGATION AT THE USER FACILITY. DURING TESTING THE DEVICE DID NOT PASS THE TOUCHSCREEN TEST. THIS WAS DUE TO THE TOUCHSCREEN BEING OUT OF CALIBRATION. THIS DEVICE HAS BEEN IDENTIFIED AS PART OF A PRODUCT RECALL. THIS REPORT REPRESENTS ALL THE INFORMATION KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.

Description of Event or Problem · 1

DURING TESTING AT THE USER FACILITY, THE TOUCHSCREEN WOULD NOT CALIBRATE. PRIOR TO TESTING, THE TOUCHSCREEN WOULD NOT CALIBRATE. PRIOR TO TESTING, THE DEVICE HAD BEEN RETURNED TO THE BIOMEDICAL DEPARTMENT FOR AN UNSPECIFIED REASON. NO INFORMATION WAS PROVIDED; THEREFORE, SPECIFIC PATIENT INFORMATION, DEVICE PROGRAMMING, OR EVENT DETAILS WERE NOT AVAILABLE. THERE WERE NO REPORTS OF ANY ADVERSE PATIENT EVENTS OR DELAYS IN CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
617548 SYMBIQ DCHANNEL 3.01 80FRN FRN HOSPIRA COSTA RICA LTD. NA NA

Patients

Seq Age Sex Outcome Treatment
1 UNK