FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 4152369 · Received October 3, 2014

Report

Report Number
1720753-2014-08394
Event Type
Malfunction
Date Received
October 3, 2014
Date of Event
September 16, 2014
Report Date
October 3, 2014
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. THE VOLTAGE ON MONITOR POWER SUPPLY WAS EVALUATED AND READJUSTED. THE FUSE WAS REPLACED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND RETURNED TO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE LEFT MONITOR WAS NOT WORKING. THIS RESULTED IN A TOTAL LOSS OF IMAGING FUNCTIONALITY. THIS COULD RESULT IN THE DELAY OR CANCELLATION OF A PROCEDURE. THERE IS NO REPORT OF INJURY OR DEATH ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
617697 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900

Patients

Seq Age Sex Outcome Treatment
1