SMALL BATTERY DRIVE
Report
- Report Number
- 8030965-2014-10546
- Event Type
- Malfunction
- Date Received
- October 8, 2014
- Date of Event
- September 24, 2014
- Report Date
- September 24, 2014
- Manufacturer
- SYNTHES OBERDORF
- Product Code
- HWE
- PMA / PMN Number
- PEXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- LA, US
- Reporter Occupation
- OTHER
Narratives
THE ACTUAL DEVICE WAS RETURNED FOR EVALUATION. RELIABILITY ENGINEERING EVALUATED THE DEVICE AND THE REPORTED CONDITION WAS DUPLICATED AND CONFIRMED. THE ASSIGNABLE ROOT CAUSE WAS DETERMINED TO BE DUE TO USAGE WEAR OVER TIME. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
IT WAS REPORTED THAT PRIOR TO SURGERY, IT WAS OBSERVED THAT THE TRIGGER ON THE SMALL BATTERY DRIVE DEVICE WAS STICKING. THERE WERE NO DELAYS TO THE SCHEDULED SURGICAL PROCEDURE AS AN IDENTICAL SPARE DEVICE WAS AVAILABLE. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 632679 | SMALL BATTERY DRIVE | INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT | HWE | SYNTHES OBERDORF | 003362 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |