FDA Adverse Event
Malfunction
Summary report: N
PORTABLE OXYGEN GENERATOR (CONCENTRATOR, HOMEFILL)
MDR report key: 4151991
·
Received October 8, 2014
Report
- Report Number
- 1031452-2014-12713
- Event Type
- Malfunction
- Date Received
- October 8, 2014
- Date of Event
- September 19, 2014
- Report Date
- September 19, 2014
- Manufacturer
- INVACARE FLORIDA OPERATIONS
- Product Code
- CAW
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AL, US
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
Description of Event or Problem · 1
PER PROVIDER, ALARMING WITH LOW O2. INDICATION IS THAT SIEVE BED FILTERS (BOTTOM) ARE NOT SEATED PROPERLY ALLOWING SIEVE MATERIAL TO BECOME PULVERIZED AND BYPASS FILTER. SIEVE DUST IS MIGRATING INTO THE MANIFOLD VALVE AND MUFFLER. THIS RECORD HAS BEEN IDENTIFIED AS PART OF CAPA (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 632624 | PORTABLE OXYGEN GENERATOR (CONCENTRATOR, HOMEFILL) | 868.5440 | CAW | INVACARE FLORIDA OPERATIONS | IRC5PO2V |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |