FDA Adverse Event Injury Summary report: N

HT50 VENTILATOR

MDR report key: 4151986 · Received July 21, 2014

Report

Report Number
2023050-2014-00313
Event Type
Injury
Date Received
July 21, 2014
Date of Event
July 4, 2014
Report Date
July 8, 2014
Manufacturer
FLIGHT MEDICAL INNOVATION
Product Code
NOU
Adverse Event
Yes
Product Problem
Yes
Report Source
Distributor report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

IT WAS REPORTED THAT A COVIDIEN SALES REPRESENTATIVE INSPECTED THE DEVICE AND NO ANOMALIES WERE FOUND. ADDITIONALLY, THE CIRCUIT AND OXYGEN BLENDING KIT WAS CHECKED, NO ANOMALIES WERE FOUND. THE DEVICE WAS REPORTED TO BE CONTINUOUSLY USED WITHOUT ISSUE. THERE WAS NO PATIENT HARM REPORTED. THE PATIENT WAS REPORTED TO REMAIN ON THE DEVICE FOR CONTINUOUS VENTILATION. NO ADDITIONAL INFORMATION WAS AVAILABLE AT THE TIME OF THE REPORT. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED BY THE USER FACILITY THAT A PATIENT ARRIVED TO THE HOSPITAL FOR RESPITE CARE BY TAXI; WHILE BEING VENTILATED ON AN HT50 VENTILATOR. THE HT50 VENTILATOR WAS REPORTED TO HAVE BEEN PLUGGED INTO THE TAXI'S LIGHTER RECEPTACLE FOR THE DURATION OF THE TRIP WHICH WAS REPORTED TO BE ABOUT AN HOUR. UPON ARRIVAL TO THE HOSPITAL THE STAFF NOTICED THE VENTILATOR WAS NOT VENTILATING THE PATIENT. THE STAFF NOTICED THE LIGHT EMITTING DIODE (LED) OF THE EXTERNAL BATTERY WAS ILLUMINATED. THE PATIENT WAS REMOVED FROM THE DEVICE AND AMBU BAGGED DURING TRANSFER FROM THE TAXI TO THE HOSPITAL. ONCE ENTERING THE HOSPITAL THE HT50 WAS REBOOTED AND PLACED BACK ON THE PATIENT. AFTER REMOVING THE PATIENT FROM THE ELEVATOR THE STAFF NOTICED THE VENTILATOR STOPPED CYCLING WITH THE ALARM LED ILLUMINATED. THE VENTILATOR CIRCUITS WERE CHECKED FOR LEAKS HOWEVER NONE WERE FOUND. IN ADDITION, THE MANOMETER WAS FOUND NOT TO BE FUNCTIONING AT THIS TIME. THE PATIENT WAS AGAIN AMBU BAGGED UNTIL A FAMILY MEMBER OF THE PATIENT REBOOTED THE VENTILATOR.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
427188 HT50 VENTILATOR CONTINUOUS VENTILATOR NOU FLIGHT MEDICAL INNOVATION HT50

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention