PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
Report
- Report Number
- 3004209178-2014-92684
- Event Type
- Malfunction
- Date Received
- October 8, 2014
- Date of Event
- July 10, 2014
- Report Date
- July 10, 2014
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KY
- Reporter Occupation
- PATIENT
Narratives
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
THE INSULIN PUMP ALARMED MOTOR ERROR DURING REWIND DUE TO CORRODED MOTOR HOME SWITCH. UNABLE TO PERFORM DISPLACEMENT TEST DUE TO MOTOR ERROR ALARM. THE INSULIN PUMP WAS RECEIVED WITH MINOR SCRATCHED DISPLAY WINDOW, CRACKED CASE AT DISPLAY WINDOW CORNERS, CRACKED BATTERY TUBE THREADS AND CRACKED RESERVOIR TUBE LIP.
IT WAS REPORTED THAT THE CUSTOMER HAD A MOTOR ERROR ALARM OCCUR WITH THEIR INSULIN PUMP. CUSTOMER'S BLOOD GLUCOSE LEVEL WAS REPORTED AT 120 MG/DL. PRODUCT IS BEING REPLACED. NO FURTHER INFORMATION PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 632691 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-723LNAL |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 42 YR |